Calibration of a DI-1000U to a load cell using a Serial Port Terminal Program
The following calibation procedures do not apply to the DI-1000HS-1K sensor interface.
Device Setup
Installing Drivers
If you have purchased LoadVUE Pro, install LoadVUE Pro. This will also install the drivers needed.
If you did not buy or do not have LoadVUE Pro, see Drivers and Serial Port Terminal Programs to download and install the drivers.
You will need to download and install a terminal program (Putty or HyperTerminal) before you can proceed further.
Connecting the Load Cell to the DI-1000
Connect the load cell wires to the DI-1000U unit to make sure the connections are correct and make sure the terminal block is inserted into the DI-1000 tightly.
Powering the DI-1000
The DI-1000U is powered by the USB and does not need external power. If you have a DI-1000Z or DI-1000ZP, then connect the power supply first.
Then connect the provided XBee USB Adapter to the computer’s USB port.
Once the DI-1000 is connected to the USB port of the PC and the drivers are properly installed, it will add a USB Serial Port entry to the Device Manager.
Connecting via Putty
The default parameters are: 9600 (or 230400 for DI-1000UHS/DI-1000UHS-1K), N,8,1,No hardware flow control.
Once connected, Press Enter in the terminal window and you should get a response ‘A’ that shows connection was successful. Now you can proceed with the rest of the procedure.
mVolt Calibration
To calibrate a load cell in pounds using mVolt/V type the following commands in sequence in the terminal window. Commands are not case sensitive. For example, to calibrate a 250 lb load cell with a 3.02 mVolt/V rating:
UNITS LB
LC 250
CAL m
MVOLT 3.02 (This number will be available from the load cell calibration/specification certificate.)
TARE (this zeros out the load)
WC (this will continuously output the load in lb. Change the load and ensure that the output changes).
To calibrate a load cell in kg using mVolt/V type the following commands in sequence in the terminal window. Commands are not case sensitive. For example, to calibrate a 250 kg load cell with a 3.02 mVolt/V rating:
LC 250
CAL m
MVOLT 3.02 (This number will be available from the load cell calibration/specification certificate.)
TARE (this zeros out the load)
WC (this will continuously output the load in kg. Change the load and ensure that the output changes).
2-Point Calibration
To calibrate a load cell in pounds using a known weight type the following commands in sequence. Commands are not case sensitive. For example, to calibrate a 250 lb load cell using a known load of 100 lb, remove all load from load cell and type:
LC 250
CAL 2
2PCAL 100
Apply Point 2 Load of 100 LB Press C when ready or Q to quit Apply 100 lb, then type C
WC (this will continuously output the load in lb. Change the load and ensure that the output changes).
To calibrate a load cell in kg using a known weight type the following commands in sequence. Commands are not case sensitive. For example, to calibrate a 250 kg load cell using a known load of 100 kg, remove all load from load cell and type:
LC 250
CAL 2
2PCAL 100
Apply Point 2 Load of 100 KG Press C when ready or Q to quit Apply 100 kg, then type C
WC (this will continuously output the load in kg. Change the load and ensure that the output changes).
To calibrate a load cell in Newtons, use UNITS N in the procedures above.
Only in some versions of the DI-1000 you can calibrate in one units (say lb) and then read the output in a different units (say Newtons) - i.e the DI-1000U performs the desired unit conversion. We recommend against doing this. Use our LoadVUE software to read the values in a different units if necessary, or, if you are writing your own software, read the values in the calibrated units and perform the unit conversion yourself.







